Viewing Study NCT05139693


Ignite Creation Date: 2025-12-25 @ 1:24 AM
Ignite Modification Date: 2025-12-25 @ 11:34 PM
Study NCT ID: NCT05139693
Status: COMPLETED
Last Update Posted: 2022-04-29
First Post: 2021-11-17
Is NOT Gene Therapy: False
Has Adverse Events: False

Brief Title: Accuracy of Pulse Oximeters With Profound Hypoxia
Sponsor: Respiree Pte Ltd
Organization:

Study Overview

Official Title: Accuracy of Pulse Oximeters With Profound Hypoxia
Status: COMPLETED
Status Verified Date: 2022-04
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: None
Brief Summary: The aim of this project is to test the accuracy of pulse oximeters during mild, moderate and severe hypoxia. This is done by comparing the reading of the pulse oximeter during brief, steady state hypoxia with a gold-standard measurement of blood oxyhemoglobin saturation (arterial blood sample processed in a laboratory hemoximeter). The data obtained is submitted by pulse oximeter manufacturers to the FDA for device approval.
Detailed Description: The study will include at least 10 subjects (up to 14 if needed to reach the 200 necessary data points to meet the ISO 80601-2-61:2017).

Per FDA guidance, at least 2, or 15% of the subjects will have dark skin.

Study Oversight

Has Oversight DMC: None
Is a FDA Regulated Drug?: False
Is a FDA Regulated Device?: True
Is an Unapproved Device?: True
Is a PPSD?: None
Is a US Export?: None
Is an FDA AA801 Violation?: