Viewing Study NCT05640323


Ignite Creation Date: 2025-12-18 @ 8:19 AM
Ignite Modification Date: 2025-12-23 @ 6:23 PM
Study NCT ID: NCT05640323
Status: None
Last Update Posted: 2023-12-20 00:00:00
First Post: 2022-11-28 00:00:00
Is Possible Gene Therapy: False
Has Adverse Events: False

Brief Title: Prospective Study to Validate the Clinical Accuracy of the Norbert Device to Measure Pulse Rate
Sponsor: None
Organization:

Study Overview

Official Title: Prospective Study to Validate the Clinical Accuracy of the Norbert Device to Measure Pulse Rate
Status: None
Status Verified Date: 2023-12
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: None
Brief Summary: The study design is to validate the pulse and heart rate accuracy as required per ISO 80601-2-61, Clause 201.12.1.104. The study is a non-randomized, non-blinded single arm design to evaluate the clinical accuracy of the ND's pulse rate measurement compared to a reference heart rate manually measured by observing R-peaks in a standard concurrently recorded ECG from an FDA 510(k)-cleared reference clinical ECG (RCECG). The sponsor will ensure that the study staff enrolls a range of types of arrhythmias including tachycardia, bradycardia, atrial fibrillation, bigeminy, trigeminy, and others.
Detailed Description: None

Study Oversight

Has Oversight DMC: None
Is a FDA Regulated Drug?: None
Is a FDA Regulated Device?: None
Is an Unapproved Device?: True
Is a PPSD?: None
Is a US Export?: None
Is an FDA AA801 Violation?: