Viewing Study NCT03678103


Ignite Creation Date: 2025-12-24 @ 7:44 PM
Ignite Modification Date: 2026-01-01 @ 6:35 AM
Study NCT ID: NCT03678103
Status: COMPLETED
Last Update Posted: 2021-09-30
First Post: 2018-09-17
Is NOT Gene Therapy: True
Has Adverse Events: True

Brief Title: Oxygen Reserve Index (ORi) Validation of INVSENSOR00029
Sponsor: Masimo Corporation
Organization:

Study Overview

Official Title: Oxygen Reserve Index (ORi) Validation of INVSENSOR00029
Status: COMPLETED
Status Verified Date: 2021-09
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: None
Brief Summary: ORi is a unitless index that could help clinicians with their assessments or normoxic and hyperoxic states by scaling the measured absorption information between 0.00 and 1.00. An ORi of 0.00 corresponds to PaO2 values of 100mmHg and below and an ORi of 1.00 corresponds to PaO2 values of 200mmHg and above. In this study, the PaO2 is varied by controlling the concentration of oxygen the study volunteer breathes. Changes in ORi values are analyzed by comparing it to corresponding changes in PaO2 measurements from blood samples.
Detailed Description: None

Study Oversight

Has Oversight DMC: None
Is a FDA Regulated Drug?: None
Is a FDA Regulated Device?: None
Is an Unapproved Device?: True
Is a PPSD?: None
Is a US Export?: None
Is an FDA AA801 Violation?: