Brief Title: Clinical Trial of H2-BP Electronic Sphygmomanometer
Sponsor: Gangnam Severance Hospital
Organization: Gangnam Severance Hospital

Study Overview

Official Title: A Prospective Single Arm Single Center Non-randomized Open Label Study to Evaluate the Accuracy of the Wrist Blood Pressure Monitor H2-BP Comparing With Aneroid Sphygmomanometer
Status: COMPLETED
Status Verified Date: 2024-10
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: None
Brief Summary: The study compares H2-BP a wristband blood pressure monitor with an auscultatory sphygmomanometer Both devices are used on the same arm After confirming wrist and upper arm circumference H2-BP is operated by one measurer while the auscultatory sphygmomanometer is simultaneously operated by two other measurers Each subject undergoes alternating measurements of H2-BP on one wrist and the auscultatory sphygmomanometer on the same arm until three valid measurement pairs are obtained per subject with a maximum of eight repetitions allowed The two auscultatory sphygmomanometer measurers operate independently and cannot see each others measurements during the test Measurements are repeated for 10 to 30 minutes until three valid blood pressure values are obtained and recorded Validity assessment variables are derived from the recorded values to evaluate the accuracy of H2-BP
Detailed Description: None

Study Oversight

Has Oversight DMC: None
Is a FDA Regulated Drug?: False
Is a FDA Regulated Device?: False
Is an Unapproved Device?: None
Is a PPSD?: None
Is a US Export?: None
Is an FDA AA801 Violation?: None