Viewing Study NCT03396003


Ignite Creation Date: 2025-12-24 @ 7:46 PM
Ignite Modification Date: 2026-01-04 @ 7:26 AM
Study NCT ID: NCT03396003
Status: COMPLETED
Last Update Posted: 2018-09-26
First Post: 2017-11-20
Is Gene Therapy: True
Has Adverse Events: False

Brief Title: A Comparison of Two Devices for Taking Measurements of the Eye to Assist in Lens Selection for Cataract Surgery
Sponsor: Ziemer Ophthalmic Systems AG
Organization:

Study Overview

Official Title: GALILEI G6 Lens Professional vs. a Predicate Device: a Comparison Study
Status: COMPLETED
Status Verified Date: 2018-09
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: None
Brief Summary: The purpose of this study is to compare the Ziemer Ophthalmic Systems AG GALILEI G6 Lens Professional to the Oculus Pentacam® AXL for taking images and measurements of the anterior segment of the eye, including the cornea, pupil, anterior chamber and lens, to assist in determining the power of the intraocular lens for implantation.
Detailed Description: A corneal topographer/biometer named GALILEI G6 Lens Professional was developed by Ziemer Ophthalmic Systems AG for the measurement of anterior segment geometry and axial intraocular distances. The measurement principle of the corneal topographer corresponds to that of the commercially available Galilei G4 Dual Scheimpflug Analyzer, whereas the measurement principle of the biometer is based on low coherence interferometry/reflectometry, which has been widely used and applied clinically by a number of predicate devices.

One complete measurement consists of a corneal topography/ tomography scan followed by three consecutive, axial biometry scans of the anterior segment (cornea and crystalline lens) and three consecutive, axial biometry scans of the retina. The measurement process is continuous but divided into three alignment-click-steps.

Such measurements can be applied to given Intra Ocular Lens (IOL) types and IOL equations, to permit the calculation of recommended IOLs to be implanted during cataract surgery in order to achieve the desired vision correction.

The primary objectives of this clinical study are to: 1) evaluate the inter-device repeatability, inter-operator repeatability and reproducibility of the GALILEI G6 Lens Professional in repeated measurements of anterior segment geometry and axial intraocular distances and, 2) demonstrate substantial equivalence through examining agreement of the GALILEI measurements to those of the predicate device, the Pentacam® AXL.

Study Oversight

Has Oversight DMC: False
Is a FDA Regulated Drug?: False
Is a FDA Regulated Device?: True
Is an Unapproved Device?: True
Is a PPSD?: None
Is a US Export?: True
Is an FDA AA801 Violation?: