Viewing Study NCT01636804


Ignite Creation Date: 2025-12-24 @ 8:03 PM
Ignite Modification Date: 2026-01-02 @ 9:10 AM
Study NCT ID: NCT01636804
Status: COMPLETED
Last Update Posted: 2013-08-09
First Post: 2012-07-05
Is NOT Gene Therapy: True
Has Adverse Events: False

Brief Title: Incidence & Predictors of Mechanical Abnormalities of QuickSite® and QuickFlex® Left Ventricular Pacing Leads
Sponsor: Dhanunjaya Lakkireddy, MD, FACC
Organization:

Study Overview

Official Title: Incidence & Predictors of Mechanical Abnormalities of QuickSite® and QuickFlex® Left Ventricular Pacing Leads: Impact on Electrophysiological Profile
Status: COMPLETED
Status Verified Date: 2013-08
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: None
Brief Summary: On April 2, 2012 the manufacturer of the Quicksite® and Quickflex® leads, St. Jude Medical, released a voluntary medical device advisory regarding these leads' potential for problems with the silicone insulation covering one end of the lead, where it meets another section of the lead which is coated with a different insulating material (polyurethane). There has been a voluntary suspension of sales of these leads.

There are approximately 179,000 of the above leads implanted worldwide with an estimated 101,000 leads currently in service (59,000 in US). There have been 39 confirmed cases of the conductor portion of the lead becoming exposed from the silicone insulation (externalization), and the rate of occurrence of this type of problem is conservatively estimated at less than one percent. However, a review done by St. Jude Medical of returned Quicksite® and Quickflex® leads and available X-ray and fluoroscopic (x-ray movie) images of patients with these leads (from 1,219 patients) revealed a higher rate of 3-4%. It is presumed that there are patients whose leads have not yet been affected by this problem, but whose leads may show signs of structural changes on x-ray or fluoroscopy. This study seeks to determine what structural changes may be occurring in these leads, what signs might predict these changes, and what impact these changes may have on the lead and its performance.
Detailed Description: None

Study Oversight

Has Oversight DMC: False
Is a FDA Regulated Drug?: None
Is a FDA Regulated Device?: None
Is an Unapproved Device?: None
Is a PPSD?: None
Is a US Export?: None
Is an FDA AA801 Violation?: