Brief Title: Safe Infusion Device in Reducing Occupational Exposure of Nurses
Sponsor: Institut de Cancérologie de la Loire
Organization:

Study Overview

Official Title: Evaluation of a Safe Infusion Device in Reducing Occupational Exposure of Nurses to Antineoplastic Drugs: a Comparative Prospective Study. CONTAMOINS-1
Status: COMPLETED
Status Verified Date: 2019-09
Last Known Status: None
Delayed Posting: No
If Stopped, Why?: Not Stopped
Has Expanded Access: False
If Expanded Access, NCT#: N/A
Has Expanded Access, NCT# Status: N/A
Acronym: CONTAMOINS-1
Brief Summary: Monocentric comparative before / after study to evaluate the efficacy of Safe Infusion Devices (SIDs) in reducing drug exposure to nurses compared to usual perfusion practices.
Detailed Description: Despite the decreasing of environmental contamination throughout the anticancer drug circuit, the administration of chemotherapies remains at risk of occupational exposure for nurses. Many medical devices aim at securing administration, but none have been scientifically evaluated to verify the actual improvement.

A monocentric comparative before / after study was carried out in an oncology day hospital in order to evaluate the efficacy of Safe Infusion Devices in reducing drug exposure compared to usual perfusion practices (neutral solvent-purged infusers). The rate of nurses' gloves contamination was estimated. The eligible protocols must contain at least one disconnection step and one of the 10 drugs screened. To avoid false negatives, each sample of gloves was contaminated with a drop of topotecan. Association between contamination and other variables was investigated using a multivariate regression analysis.

The usual practice leaded to a rate of 58.3% of contaminated samples while Safe Infusion Devices to a rate of 15%: Safe Infusion Devices reduced the risk of gloves contamination by 84% in multivariate analysis (Odds ratio=0.16; 95% confidence interval=0.05-0.47; p\<0.001). Topotecan was identified within 100% of the samples. Only one case of cross-contamination has occurred.

Despite the current practice of using neutral solvent-purged infusers, the occupational exposure remains high for nurses and Safe Infusion Devices significantly reduced this risk of exposure. However, glove contamination is only a surrogate endpoint. The results confirmed that the disconnection of empty bags resulted in occupational exposure. Except a contamination due to the leakage of a bag, no cross-contamination was detected. This validated the environmental quality of the cytotoxic drug circuit. Safe Infusion Devices were highly effective but did not completely eliminate exposure.

Study Oversight

Has Oversight DMC: False
Is a FDA Regulated Drug?: False
Is a FDA Regulated Device?: False
Is an Unapproved Device?: None
Is a PPSD?: None
Is a US Export?: None
Is an FDA AA801 Violation?:

Secondary ID Infos

Secondary ID Type Domain Link View
IRBN522017/CHUSTE OTHER Comité d'Ethique du CHU de Saint-Etienne View